中国地质学会岩矿测试技术专业委员会、国家地质实验测试中心主办

193 nm ArF准分子激光剥蚀等离子体质谱法测定熔融玻璃中微量铌和钽

王建其, 林慈銮, 袁洪林, 柳小明, 第五春荣. 193 nm ArF准分子激光剥蚀等离子体质谱法测定熔融玻璃中微量铌和钽[J]. 岩矿测试, 2007, 26(1): 1-3.
引用本文: 王建其, 林慈銮, 袁洪林, 柳小明, 第五春荣. 193 nm ArF准分子激光剥蚀等离子体质谱法测定熔融玻璃中微量铌和钽[J]. 岩矿测试, 2007, 26(1): 1-3.
Determination of Micro-amount of Niobium and Tantalum in Fused Glass Samples by Argon Fluoride 193 nm Excimer Laser Ablation-Inductively Coupled Plasma Mass Spectrometry[J]. Rock and Mineral Analysis, 2007, 26(1): 1-3.
Citation: Determination of Micro-amount of Niobium and Tantalum in Fused Glass Samples by Argon Fluoride 193 nm Excimer Laser Ablation-Inductively Coupled Plasma Mass Spectrometry[J]. Rock and Mineral Analysis, 2007, 26(1): 1-3.

193 nm ArF准分子激光剥蚀等离子体质谱法测定熔融玻璃中微量铌和钽

  • 基金项目:

    教育部科学技术研究重点项目

    国家自然科学基金

Determination of Micro-amount of Niobium and Tantalum in Fused Glass Samples by Argon Fluoride 193 nm Excimer Laser Ablation-Inductively Coupled Plasma Mass Spectrometry

  • 建立了一种利用193 nm ArF准分子激光剥蚀-等离子体质谱法直接测定岩石熔融玻璃中Nb和Ta的分析方法。采用常规的碱熔法熔融样品,以29Si为内标、玻璃标准参考物质NIST SRM612为外标进行测定,方法的检出限分别为Nb0.003μg/g和Ta0.006μg/g(n=10)。分析4个不同的地质标准参考物质结果表明,Nb和Ta测定结果的相对误差(RE)和精密度(RSD,n=11)均低于10%。
  • 加载中
  • 加载中
计量
  • 文章访问数:  1399
  • PDF下载数:  1
  • 施引文献:  0
出版历程
修回日期:  2006-05-14
刊出日期:  2007-02-28

目录