中国地质学会岩矿测试技术专业委员会、国家地质实验测试中心主办

二次熔矿方式电感耦合等离子体发射光谱法测定棕刚玉中硅铁钛钙镁锆

杨载明. 二次熔矿方式电感耦合等离子体发射光谱法测定棕刚玉中硅铁钛钙镁锆[J]. 岩矿测试, 2012, 31(4): 617-620.
引用本文: 杨载明. 二次熔矿方式电感耦合等离子体发射光谱法测定棕刚玉中硅铁钛钙镁锆[J]. 岩矿测试, 2012, 31(4): 617-620.
YANG Zai-ming. Simultaneous Quantification Measurement of Si, Fe, Ti, Ca, Mg, Zr in Corundum by Inductively Coupled Plasma-Atomic Emission Spectrometry with Secondary-Melting Method[J]. Rock and Mineral Analysis, 2012, 31(4): 617-620.
Citation: YANG Zai-ming. Simultaneous Quantification Measurement of Si, Fe, Ti, Ca, Mg, Zr in Corundum by Inductively Coupled Plasma-Atomic Emission Spectrometry with Secondary-Melting Method[J]. Rock and Mineral Analysis, 2012, 31(4): 617-620.

二次熔矿方式电感耦合等离子体发射光谱法测定棕刚玉中硅铁钛钙镁锆

Simultaneous Quantification Measurement of Si, Fe, Ti, Ca, Mg, Zr in Corundum by Inductively Coupled Plasma-Atomic Emission Spectrometry with Secondary-Melting Method

  • 根据棕刚玉难熔的特点,采用二次熔矿方式对样品进行分解,经盐酸酸化的浸提液用电感耦合等离子体发射光谱法测定硅、铁、钛、钙、镁、锆的含量。第一次加入硼酸-碳酸钠(质量比2∶1)熔矿,熔剂中大量硼酸的存在可以确保棕刚玉样品无需研磨而分解完全;但所得熔融物极难提取,通过提高熔剂中碳酸钠的比例后进行第二次熔矿,使熔融物酸化浸取变得快速简单。对硼酸-碳酸钠混合溶剂的比例、熔矿温度和时间、称样量和样品粒度等实验条件进行优化,分析结果表明方法检出限为0.0003%~0.0051%,精密度(RSD,n=11)为0.88%~4.78%。用标准物质验证,测定值与标准值相符;与国标法对比验证,测定结果相符。该方法样品分解充分,对样品的粒度适应范围宽,可满足生产控制快速检测的要求。
  • 加载中
  • 加载中
计量
  • 文章访问数:  1457
  • PDF下载数:  1
  • 施引文献:  0
出版历程
收稿日期:  2011-11-25
刊出日期:  2012-08-28

目录