Determination of Free α-SiO2 Content in Mill Scale by X-ray Diffraction K Value Method
-
摘要:
目前通常采用焦磷酸法和X射线衍射法测定游离α-SiO2的含量, 其中焦磷酸法不能消除氧化铁皮中FeO、Fe3O4和Fe2O3等焦磷酸难溶物质的影响, 不适合用于测定氧化铁皮中的游离α-SiO2。本文建立了采用X射线衍射K值法测定氧化铁皮中游离α-SiO2含量的方法。以α-Al2O3作为参考物质, 以α-SiO2(101) 衍射峰和α-Al2O3(012) 衍射峰作为测量谱峰, 将α-SiO2和α-Al2O3(质量比1:1) 混合均匀制备成测量K值的试样, 获得K值为7.86, 应用此值分析已知游离α-SiO2含量的氧化铁皮样品, 游离α-SiO2含量的测定值与实际值相符。该方法的测定范围为游离α-SiO2含量≥0.50%, 测量范围宽, 且方便快速, 能够满足进口氧化铁皮检验监管的工作需求。
-
关键词:
- 氧化铁皮 /
- 游离α-SiO2含量 /
- X射线衍射法 /
- K值法
Abstract:At present, free α-SiO2 content is often determined by the pyrophosphate method and X-ray Diffraction (XRD) method. However, insoluble materials including FeO, Fe3O4 and Fe2O3 in mill scale cannot be decomposed by pyrophosphate, and thus the pyrophosphate method is not suitable for the determination of free α-SiO2 content in mill scale. A method for detecting free α-SiO2 content in mill scale through X-ray diffraction K value method has been developed. α-Al2O3 was used as the reference material. Diffraction peaks of crystal face (101) for α-SiO2 and crystal face (012) for α-Al2O3 were used for measured diffraction peaks. Samples for detecting K value were prepared by mixing α-SiO2 with α-Al2O3 at a mass ratio of 1:1. The acquired K value is 7.86. The measured K value has been applied to the determination of mill scale samples with known free α-SiO2 content, and the analytical results of free α-SiO2 content are in good agreement with practical values. Detection range of this method is that free α-SiO2 are equal or greater than 0.5%. This method is convenient and rapid, and has a wide detection range and can meet the need of imported mill scale inspection and supervision work.
-
Key words:
- mill scale /
- free α-SiO2 content /
- X-ray Diffraction /
- K value method
-
-
表 1 衍射峰积分强度测量结果
Table 1. Measurement results of diffraction intensity
样品
编号α-SiO2(101) 衍射峰积分强度 α-Al2O3(012) 衍射峰积分强度 第1次
测量值
(cps)第2次
测量值
(cps)相对平均
偏差
(%)第1次
测量值
(cps)第2次
测量值
(cps)相对平均
偏差
(%)A 68.97 69.00 0.022 8.782 8.781 0.0057 B 0.7033 0.6949 0.60 0.7335 0.7324 0.075 表 2 K值测试结果
Table 2. Measurement results of K value
制样
序次测量
序次衍射峰积分强度(cps) 单次测量的
K值K的
平均值α-SiO2
(101)α-Al2O3
(012)1 1 68.97 8.782 7.85 7.86 2 69.00 8.781 7.86 3 68.95 8.811 7.83 2 1 70.29 8.882 7.91 2 69.99 8.847 7.91 3 70.01 8.874 7.89 3 1 69.11 8.795 7.86 2 69.14 8.855 7.81 3 69.26 8.841 7.83 表 3 样品配比及游离α-SiO2含量测定结果
Table 3. Proportioning of samples and determination results of free α-SiO2 content
样品
编号样品配比 α-Al2O3
加入量(g)衍射峰积分强度(cps) α-SiO2含量(%) 相对偏差
(%)1#氧化铁皮质量(g) 石英砂质量(g) α-SiO2(101) α-Al2O3(012) 测定值 实际值 2# 2 0.002 0.016 - - - 0.10 - 3# 2 0.01 0.08 0.1685 0.1631 0.52 0.50 4.00 4# 2 0.02 0.16 0.3099 0.3001 1.05 1.00 5.00 5# 2 0.05 0.4 0.7033 0.7335 2.38 2.44 2.46 6# 2 0.07 0.56 0.9614 0.9713 3.40 3.38 0.59 -
[1] [2] [3] [4] [5] [6] [7] [8] [9] [10] [11] [12] [13] [14] [15] -